Bug #7438
DNA PCR requisition for PID 0502-2-60 at 2000D
Status: | Feedback | Start date: | 01/03/2024 | |
---|---|---|---|---|
Priority: | High | Due date: | 06/03/2024 | |
Assignee: | Ame Diphoko | % Done: | 0% | |
Category: | - | Spent time: | - | |
Target version: | - |
Description
This PID has a DNA PCR requisition already existing on Senaite.I wanted to capture this requisition on the EDC however, the CRF throws an error(but the sample id matches the one on senaite)-see as attached.
History
#1 Updated by Ame Diphoko 11 months ago
- Due date set to 06/03/2024
- Status changed from New to Resolved
- Assignee set to Ame Diphoko
Requisition was throwing a validation due to the sample type WB-EDTA not matching EDC configured type DBS for DNA PCR. We have added the requisition manually on the backend to avoid the validation since sample has already been sent to lab for testing.
1. We will explore if it's possible to add a validation on the LIMS to restrict DNA PCR to only DBS type. The configuration type on EDC remains as DBS and we are expected to use the same on the LIS.
#2 Updated by Gosego Masasa 11 months ago
- Status changed from Resolved to Feedback
Thanks Ame. Will leave this in feedback to remind you to explore the possibility of adding a validation on the Senaite system to restrict DNA PCR to only DBS type as sample type.